Computer‐Aided Diagnosis in the Evaluation of Thyroid Nodules: A Study of Intra‐ And Inter‐Rater Reliability and Agreement
Head & Neck, 2026
Abstract
Objective: To evaluate the intra‐ and inter‐rater reliability of a computer‐aided diagnosis system applied to thyroid nodule assessment.
Methods: This prospective, single‐center study included 150 thyroid nodules evaluated by two physicians at two time points, 90 days apart. Analyses were performed using the AmCAD‐UT system, focusing on morphological features and ACR TI‐RADS classification. Cohen's kappa coefficient and percentage agreement were used to assess reliability.
Results: Intra‐rater reliability ranged from moderate to almost perfect, with kappa values from 0.49 (95% CI: 0.31–0.66) to 0.98 (95% CI: 0.96–1.00), and agreement rates from 81.3% to 99.3%. Rater 2 demonstrated higher reproducibility across most variables, particularly for "texture" (k = 0.98), "margin" (k = 0.90), "composition" (k = 0.93), and "taller‐than‐wide" (k = 0.92). Inter‐rater agreement was more variable, with kappa values ranging from 0.43 (95% CI: 0.23–0.62) to 0.96 (95% CI: 0.89–1.00), and agreement percentages from 78.0% to 99.3%. The lowest inter‐rater reproducibility was observed for "shape".
Conclusion: The computer‐aided diagnosis system demonstrated predominantly moderate to almost perfect intra‐rater reliability and moderate to strong inter‐rater agreement across most evaluated features. The highest reproducibility was observed for "taller‐than‐wide," "texture," and "composition," whereas "shape" consistently showed lower agreement. These findings support the system's role as a reliable adjunct for standardizing thyroid nodule assessment, although its performance remains partially influenced by operational factors and warrants further multicenter validation.